Cargando…

T wave oversensing in subcutaneous implantable cardioverter defibrillator secondary to hematoma formation: A potential cause of early postimplantation inappropriate shocks

T wave oversensing (TWOS) is the most common cause of inappropriate shocks in subcutaneous cardioverter‐defibrillators (S‐ICD). We are presenting a patient with severe ischemic cardiomyopathy who received a S‐ICD while on antiplatelets therapy. Pressure dressing was applied due to significant bleedi...

Descripción completa

Detalles Bibliográficos
Autores principales: Halawa, Ahmad, Gautam, Sandeep
Formato: Online Artículo Texto
Lenguaje:English
Publicado: John Wiley and Sons Inc. 2018
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC6373653/
https://www.ncbi.nlm.nih.gov/pubmed/30805053
http://dx.doi.org/10.1002/joa3.12132