Inappropriate implantable cardioverter defibrillator shocks due to atrial far‐field on the tip‐to‐ring channel for lead dislodgement

We reported a case of inappropriate implantable cardioverter defibrillator shocks, due to atrial far‐field on the tip‐to‐ring channel of the fast electrical activity during atrial fibrillation, caused by lead dislocation in the right ventricle outflow tract. During these episodes the can‐to‐right ve...