Customisation and validation of a low-volume plasma renin activity immunoassay: Enabling of regulatory compliant determination in paediatric trials

OBJECTIVES: Investigations of plasma renin activity (PRA) in children are urgently required. Small-volume, regulatory guideline compliant, bioanalytical assays tailored for paediatric application could facilitate to overcome this hurdle. Ethical constraints given e.g. by the European Medicines Agenc...

Descripción completa

Detalles Bibliográficos
Autores principales: Suessenbach, F.K., Tins, J., Burckhardt, B.B.
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Elsevier 2019
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC6904673/
https://www.ncbi.nlm.nih.gov/pubmed/31867426
http://dx.doi.org/10.1016/j.plabm.2019.e00144
_version_ 1783478038870097920
author Suessenbach, F.K.
Tins, J.
Burckhardt, B.B.
author_facet Suessenbach, F.K.
Tins, J.
Burckhardt, B.B.
author_sort Suessenbach, F.K.
collection PubMed
description OBJECTIVES: Investigations of plasma renin activity (PRA) in children are urgently required. Small-volume, regulatory guideline compliant, bioanalytical assays tailored for paediatric application could facilitate to overcome this hurdle. Ethical constraints given e.g. by the European Medicines Agency need to be addressed and reliable data generation in line with Good Clinical Laboratory Practice must be ensured. METHODS: A PRA enzyme-linked immunosorbent assay (ELISA) was tailored for paediatric application and validated in the context of the U.S. Food and Drug Administration bioanalytical guideline. Performance verification of the assay was conducted by participation in an interlaboratory ring test, evaluation of incurred sample reanalysis and an application-orientated approach in children. RESULTS: A five-fold reduction of required plasma volume to 100 μL was achieved without limiting the calibration range. Between-run accuracy and precision varied no more than 5.0% and 6.3%, respectively. No substantial matrix effect was detected and the inter-run precision for parallelism was 11.1%. Stability experiments approved the freeze-thaw stability, short-term stability as well as 37 weeks of long-term stability. The assay successfully participated in the interlaboratory ring test, showing non-inferiority regarding radioimmunoassay (RIA). Moreover, PRA in plasma samples of neonates was successfully determined. Conducted incurred sample reanalysis confirmed the comparability and reliability of the assay with regard to international regulatory bioanalytical guidelines. CONCLUSION: A fit-for-purpose PRA ELISA characterised by low-volume application was successfully established, indicating non-inferiority regarding commonly applied RIAs. Reliability of the regulatory-compliant PRA assay was proven by participation in an interlaboratory ring test and its application in a paediatric population.
format Online
Article
Text
id pubmed-6904673
institution National Center for Biotechnology Information
language English
publishDate 2019
publisher Elsevier
record_format MEDLINE/PubMed
spelling pubmed-69046732019-12-20 Customisation and validation of a low-volume plasma renin activity immunoassay: Enabling of regulatory compliant determination in paediatric trials Suessenbach, F.K. Tins, J. Burckhardt, B.B. Pract Lab Med Article OBJECTIVES: Investigations of plasma renin activity (PRA) in children are urgently required. Small-volume, regulatory guideline compliant, bioanalytical assays tailored for paediatric application could facilitate to overcome this hurdle. Ethical constraints given e.g. by the European Medicines Agency need to be addressed and reliable data generation in line with Good Clinical Laboratory Practice must be ensured. METHODS: A PRA enzyme-linked immunosorbent assay (ELISA) was tailored for paediatric application and validated in the context of the U.S. Food and Drug Administration bioanalytical guideline. Performance verification of the assay was conducted by participation in an interlaboratory ring test, evaluation of incurred sample reanalysis and an application-orientated approach in children. RESULTS: A five-fold reduction of required plasma volume to 100 μL was achieved without limiting the calibration range. Between-run accuracy and precision varied no more than 5.0% and 6.3%, respectively. No substantial matrix effect was detected and the inter-run precision for parallelism was 11.1%. Stability experiments approved the freeze-thaw stability, short-term stability as well as 37 weeks of long-term stability. The assay successfully participated in the interlaboratory ring test, showing non-inferiority regarding radioimmunoassay (RIA). Moreover, PRA in plasma samples of neonates was successfully determined. Conducted incurred sample reanalysis confirmed the comparability and reliability of the assay with regard to international regulatory bioanalytical guidelines. CONCLUSION: A fit-for-purpose PRA ELISA characterised by low-volume application was successfully established, indicating non-inferiority regarding commonly applied RIAs. Reliability of the regulatory-compliant PRA assay was proven by participation in an interlaboratory ring test and its application in a paediatric population. Elsevier 2019-11-12 /pmc/articles/PMC6904673/ /pubmed/31867426 http://dx.doi.org/10.1016/j.plabm.2019.e00144 Text en © 2019 The Authors http://creativecommons.org/licenses/by/4.0/ This is an open access article under the CC BY license (http://creativecommons.org/licenses/by/4.0/).
spellingShingle Article
Suessenbach, F.K.
Tins, J.
Burckhardt, B.B.
Customisation and validation of a low-volume plasma renin activity immunoassay: Enabling of regulatory compliant determination in paediatric trials
title Customisation and validation of a low-volume plasma renin activity immunoassay: Enabling of regulatory compliant determination in paediatric trials
title_full Customisation and validation of a low-volume plasma renin activity immunoassay: Enabling of regulatory compliant determination in paediatric trials
title_fullStr Customisation and validation of a low-volume plasma renin activity immunoassay: Enabling of regulatory compliant determination in paediatric trials
title_full_unstemmed Customisation and validation of a low-volume plasma renin activity immunoassay: Enabling of regulatory compliant determination in paediatric trials
title_short Customisation and validation of a low-volume plasma renin activity immunoassay: Enabling of regulatory compliant determination in paediatric trials
title_sort customisation and validation of a low-volume plasma renin activity immunoassay: enabling of regulatory compliant determination in paediatric trials
topic Article
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC6904673/
https://www.ncbi.nlm.nih.gov/pubmed/31867426
http://dx.doi.org/10.1016/j.plabm.2019.e00144
work_keys_str_mv AT suessenbachfk customisationandvalidationofalowvolumeplasmareninactivityimmunoassayenablingofregulatorycompliantdeterminationinpaediatrictrials
AT tinsj customisationandvalidationofalowvolumeplasmareninactivityimmunoassayenablingofregulatorycompliantdeterminationinpaediatrictrials
AT burckhardtbb customisationandvalidationofalowvolumeplasmareninactivityimmunoassayenablingofregulatorycompliantdeterminationinpaediatrictrials
AT customisationandvalidationofalowvolumeplasmareninactivityimmunoassayenablingofregulatorycompliantdeterminationinpaediatrictrials