Twiddler’s syndrome with a subcutaneous implantable cardioverter-defibrillator presenting with an inappropriate shock: a case report

BACKGROUND: Subcutaneous implantable cardioverter-defibrillators (S-ICDs) are increasingly used in patients at risk of fatal cardiac arrhythmias. Twiddler’s syndrome is a condition in which a device is manipulated by the patient after implantation leading to lead twisting and retraction. Device mani...

Descripción completa

Detalles Bibliográficos
Autores principales: Camm, C Fielder, Rajappan, Kim, Curson, Mark, Tilling, Lindsey
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Oxford University Press 2019
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC7042151/
https://www.ncbi.nlm.nih.gov/pubmed/32123805
http://dx.doi.org/10.1093/ehjcr/ytz200