Fracture of an epicardial left ventricular lead implanted at open‐heart surgery in anticipation of future need for cardiac resynchronization therapy
Epicardial left ventricular leads can be implanted at open‐heart surgery for cardiac resynchronization therapy. We report a 2‐year‐old fractured epicardial left ventricular lead detected at generator implant. It highlights the importance of good surgical implant technique and of rigorous lead evalua...