Challenges in Removing the Essure(®) Device

We present a case about Essure(®) removal surgery in which the third markers of the device have torn off. The woman needed a second surgery for complete removal of the devices. Fluoroscopy during surgery is a good method to visualize the lost fragments. With fluoroscopy, a hysterectomy is not needed...