Biochemical, immunochemical and serology analytes validation of the lithium heparin BD Barricor blood collection tube on a highly automated Roche COBAS8000 instrument
BACKGROUND: Recently developed blood tubes with a barrier to provide plasma are becoming widespread. We compared 43 biochemical, 35 immunochemical and 7 serology analytes in a BD-Vacutainer® Barricor tube for local clinical validation of this lithium-heparin tube with a barrier. METHODS: Samples fro...