Biochemical, immunochemical and serology analytes validation of the lithium heparin BD Barricor blood collection tube on a highly automated Roche COBAS8000 instrument

BACKGROUND: Recently developed blood tubes with a barrier to provide plasma are becoming widespread. We compared 43 biochemical, 35 immunochemical and 7 serology analytes in a BD-Vacutainer® Barricor tube for local clinical validation of this lithium-heparin tube with a barrier. METHODS: Samples fro...

Descripción completa

Detalles Bibliográficos
Autores principales: Ferrari, Davide, Strollo, Marta, Vidali, Matteo, Motta, Andrea, Pontillo, Marina, Locatelli, Massimo
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Mattioli 1885 2020
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC7569594/
https://www.ncbi.nlm.nih.gov/pubmed/32191654
http://dx.doi.org/10.23750/abm.v91i1.9195
_version_ 1783596760485068800
author Ferrari, Davide
Strollo, Marta
Vidali, Matteo
Motta, Andrea
Pontillo, Marina
Locatelli, Massimo
author_facet Ferrari, Davide
Strollo, Marta
Vidali, Matteo
Motta, Andrea
Pontillo, Marina
Locatelli, Massimo
author_sort Ferrari, Davide
collection PubMed
description BACKGROUND: Recently developed blood tubes with a barrier to provide plasma are becoming widespread. We compared 43 biochemical, 35 immunochemical and 7 serology analytes in a BD-Vacutainer® Barricor tube for local clinical validation of this lithium-heparin tube with a barrier. METHODS: Samples from 70 volunteers were collected in different BD-tubes: a clot-activator tube with gel (SST), a lithium-heparin tube with gel (PST), and a lithium-heparin tube with barrier (BAR). Biases from Bland-Altman plots and 95% confidence intervals were compared with the desirable specification from the Ricos database in order to verify whether measurements from different tubes were significantly different. RESULTS: For most of the analytes tested, the measurements using SST, PST or BAR tubes were equivalent. Only BIC, GLU, K, LAD, LPA, P, TP, CTX, Ferritin, HGH, vitD3 and ANTIS showed statistically significant, between-tubes, differences which might have clinical implication. CONCLUSIONS: The study demonstrates that SST, PST and BAR can be used interchangeably for most of the analytes tested, including serology analytes. This allows the use of the same tube for assaying multiple analytes, increasing the laboratory efficiency while decreasing patients discomfort by minimizing blood withdrawal. (www.actabiomedica.it)
format Online
Article
Text
id pubmed-7569594
institution National Center for Biotechnology Information
language English
publishDate 2020
publisher Mattioli 1885
record_format MEDLINE/PubMed
spelling pubmed-75695942020-10-21 Biochemical, immunochemical and serology analytes validation of the lithium heparin BD Barricor blood collection tube on a highly automated Roche COBAS8000 instrument Ferrari, Davide Strollo, Marta Vidali, Matteo Motta, Andrea Pontillo, Marina Locatelli, Massimo Acta Biomed Original Article BACKGROUND: Recently developed blood tubes with a barrier to provide plasma are becoming widespread. We compared 43 biochemical, 35 immunochemical and 7 serology analytes in a BD-Vacutainer® Barricor tube for local clinical validation of this lithium-heparin tube with a barrier. METHODS: Samples from 70 volunteers were collected in different BD-tubes: a clot-activator tube with gel (SST), a lithium-heparin tube with gel (PST), and a lithium-heparin tube with barrier (BAR). Biases from Bland-Altman plots and 95% confidence intervals were compared with the desirable specification from the Ricos database in order to verify whether measurements from different tubes were significantly different. RESULTS: For most of the analytes tested, the measurements using SST, PST or BAR tubes were equivalent. Only BIC, GLU, K, LAD, LPA, P, TP, CTX, Ferritin, HGH, vitD3 and ANTIS showed statistically significant, between-tubes, differences which might have clinical implication. CONCLUSIONS: The study demonstrates that SST, PST and BAR can be used interchangeably for most of the analytes tested, including serology analytes. This allows the use of the same tube for assaying multiple analytes, increasing the laboratory efficiency while decreasing patients discomfort by minimizing blood withdrawal. (www.actabiomedica.it) Mattioli 1885 2020 2020-03-19 /pmc/articles/PMC7569594/ /pubmed/32191654 http://dx.doi.org/10.23750/abm.v91i1.9195 Text en Copyright: © 2020 ACTA BIO MEDICA SOCIETY OF MEDICINE AND NATURAL SCIENCES OF PARMA http://creativecommons.org/licenses/by-nc-sa/4.0 This work is licensed under a Creative Commons Attribution 4.0 International License
spellingShingle Original Article
Ferrari, Davide
Strollo, Marta
Vidali, Matteo
Motta, Andrea
Pontillo, Marina
Locatelli, Massimo
Biochemical, immunochemical and serology analytes validation of the lithium heparin BD Barricor blood collection tube on a highly automated Roche COBAS8000 instrument
title Biochemical, immunochemical and serology analytes validation of the lithium heparin BD Barricor blood collection tube on a highly automated Roche COBAS8000 instrument
title_full Biochemical, immunochemical and serology analytes validation of the lithium heparin BD Barricor blood collection tube on a highly automated Roche COBAS8000 instrument
title_fullStr Biochemical, immunochemical and serology analytes validation of the lithium heparin BD Barricor blood collection tube on a highly automated Roche COBAS8000 instrument
title_full_unstemmed Biochemical, immunochemical and serology analytes validation of the lithium heparin BD Barricor blood collection tube on a highly automated Roche COBAS8000 instrument
title_short Biochemical, immunochemical and serology analytes validation of the lithium heparin BD Barricor blood collection tube on a highly automated Roche COBAS8000 instrument
title_sort biochemical, immunochemical and serology analytes validation of the lithium heparin bd barricor blood collection tube on a highly automated roche cobas8000 instrument
topic Original Article
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC7569594/
https://www.ncbi.nlm.nih.gov/pubmed/32191654
http://dx.doi.org/10.23750/abm.v91i1.9195
work_keys_str_mv AT ferraridavide biochemicalimmunochemicalandserologyanalytesvalidationofthelithiumheparinbdbarricorbloodcollectiontubeonahighlyautomatedrochecobas8000instrument
AT strollomarta biochemicalimmunochemicalandserologyanalytesvalidationofthelithiumheparinbdbarricorbloodcollectiontubeonahighlyautomatedrochecobas8000instrument
AT vidalimatteo biochemicalimmunochemicalandserologyanalytesvalidationofthelithiumheparinbdbarricorbloodcollectiontubeonahighlyautomatedrochecobas8000instrument
AT mottaandrea biochemicalimmunochemicalandserologyanalytesvalidationofthelithiumheparinbdbarricorbloodcollectiontubeonahighlyautomatedrochecobas8000instrument
AT pontillomarina biochemicalimmunochemicalandserologyanalytesvalidationofthelithiumheparinbdbarricorbloodcollectiontubeonahighlyautomatedrochecobas8000instrument
AT locatellimassimo biochemicalimmunochemicalandserologyanalytesvalidationofthelithiumheparinbdbarricorbloodcollectiontubeonahighlyautomatedrochecobas8000instrument