Biochemical, immunochemical and serology analytes validation of the lithium heparin BD Barricor blood collection tube on a highly automated Roche COBAS8000 instrument
BACKGROUND: Recently developed blood tubes with a barrier to provide plasma are becoming widespread. We compared 43 biochemical, 35 immunochemical and 7 serology analytes in a BD-Vacutainer® Barricor tube for local clinical validation of this lithium-heparin tube with a barrier. METHODS: Samples fro...
Autores principales: | , , , , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
Mattioli 1885
2020
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC7569594/ https://www.ncbi.nlm.nih.gov/pubmed/32191654 http://dx.doi.org/10.23750/abm.v91i1.9195 |
_version_ | 1783596760485068800 |
---|---|
author | Ferrari, Davide Strollo, Marta Vidali, Matteo Motta, Andrea Pontillo, Marina Locatelli, Massimo |
author_facet | Ferrari, Davide Strollo, Marta Vidali, Matteo Motta, Andrea Pontillo, Marina Locatelli, Massimo |
author_sort | Ferrari, Davide |
collection | PubMed |
description | BACKGROUND: Recently developed blood tubes with a barrier to provide plasma are becoming widespread. We compared 43 biochemical, 35 immunochemical and 7 serology analytes in a BD-Vacutainer® Barricor tube for local clinical validation of this lithium-heparin tube with a barrier. METHODS: Samples from 70 volunteers were collected in different BD-tubes: a clot-activator tube with gel (SST), a lithium-heparin tube with gel (PST), and a lithium-heparin tube with barrier (BAR). Biases from Bland-Altman plots and 95% confidence intervals were compared with the desirable specification from the Ricos database in order to verify whether measurements from different tubes were significantly different. RESULTS: For most of the analytes tested, the measurements using SST, PST or BAR tubes were equivalent. Only BIC, GLU, K, LAD, LPA, P, TP, CTX, Ferritin, HGH, vitD3 and ANTIS showed statistically significant, between-tubes, differences which might have clinical implication. CONCLUSIONS: The study demonstrates that SST, PST and BAR can be used interchangeably for most of the analytes tested, including serology analytes. This allows the use of the same tube for assaying multiple analytes, increasing the laboratory efficiency while decreasing patients discomfort by minimizing blood withdrawal. (www.actabiomedica.it) |
format | Online Article Text |
id | pubmed-7569594 |
institution | National Center for Biotechnology Information |
language | English |
publishDate | 2020 |
publisher | Mattioli 1885 |
record_format | MEDLINE/PubMed |
spelling | pubmed-75695942020-10-21 Biochemical, immunochemical and serology analytes validation of the lithium heparin BD Barricor blood collection tube on a highly automated Roche COBAS8000 instrument Ferrari, Davide Strollo, Marta Vidali, Matteo Motta, Andrea Pontillo, Marina Locatelli, Massimo Acta Biomed Original Article BACKGROUND: Recently developed blood tubes with a barrier to provide plasma are becoming widespread. We compared 43 biochemical, 35 immunochemical and 7 serology analytes in a BD-Vacutainer® Barricor tube for local clinical validation of this lithium-heparin tube with a barrier. METHODS: Samples from 70 volunteers were collected in different BD-tubes: a clot-activator tube with gel (SST), a lithium-heparin tube with gel (PST), and a lithium-heparin tube with barrier (BAR). Biases from Bland-Altman plots and 95% confidence intervals were compared with the desirable specification from the Ricos database in order to verify whether measurements from different tubes were significantly different. RESULTS: For most of the analytes tested, the measurements using SST, PST or BAR tubes were equivalent. Only BIC, GLU, K, LAD, LPA, P, TP, CTX, Ferritin, HGH, vitD3 and ANTIS showed statistically significant, between-tubes, differences which might have clinical implication. CONCLUSIONS: The study demonstrates that SST, PST and BAR can be used interchangeably for most of the analytes tested, including serology analytes. This allows the use of the same tube for assaying multiple analytes, increasing the laboratory efficiency while decreasing patients discomfort by minimizing blood withdrawal. (www.actabiomedica.it) Mattioli 1885 2020 2020-03-19 /pmc/articles/PMC7569594/ /pubmed/32191654 http://dx.doi.org/10.23750/abm.v91i1.9195 Text en Copyright: © 2020 ACTA BIO MEDICA SOCIETY OF MEDICINE AND NATURAL SCIENCES OF PARMA http://creativecommons.org/licenses/by-nc-sa/4.0 This work is licensed under a Creative Commons Attribution 4.0 International License |
spellingShingle | Original Article Ferrari, Davide Strollo, Marta Vidali, Matteo Motta, Andrea Pontillo, Marina Locatelli, Massimo Biochemical, immunochemical and serology analytes validation of the lithium heparin BD Barricor blood collection tube on a highly automated Roche COBAS8000 instrument |
title | Biochemical, immunochemical and serology analytes validation of the lithium heparin BD Barricor blood collection tube on a highly automated Roche COBAS8000 instrument |
title_full | Biochemical, immunochemical and serology analytes validation of the lithium heparin BD Barricor blood collection tube on a highly automated Roche COBAS8000 instrument |
title_fullStr | Biochemical, immunochemical and serology analytes validation of the lithium heparin BD Barricor blood collection tube on a highly automated Roche COBAS8000 instrument |
title_full_unstemmed | Biochemical, immunochemical and serology analytes validation of the lithium heparin BD Barricor blood collection tube on a highly automated Roche COBAS8000 instrument |
title_short | Biochemical, immunochemical and serology analytes validation of the lithium heparin BD Barricor blood collection tube on a highly automated Roche COBAS8000 instrument |
title_sort | biochemical, immunochemical and serology analytes validation of the lithium heparin bd barricor blood collection tube on a highly automated roche cobas8000 instrument |
topic | Original Article |
url | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC7569594/ https://www.ncbi.nlm.nih.gov/pubmed/32191654 http://dx.doi.org/10.23750/abm.v91i1.9195 |
work_keys_str_mv | AT ferraridavide biochemicalimmunochemicalandserologyanalytesvalidationofthelithiumheparinbdbarricorbloodcollectiontubeonahighlyautomatedrochecobas8000instrument AT strollomarta biochemicalimmunochemicalandserologyanalytesvalidationofthelithiumheparinbdbarricorbloodcollectiontubeonahighlyautomatedrochecobas8000instrument AT vidalimatteo biochemicalimmunochemicalandserologyanalytesvalidationofthelithiumheparinbdbarricorbloodcollectiontubeonahighlyautomatedrochecobas8000instrument AT mottaandrea biochemicalimmunochemicalandserologyanalytesvalidationofthelithiumheparinbdbarricorbloodcollectiontubeonahighlyautomatedrochecobas8000instrument AT pontillomarina biochemicalimmunochemicalandserologyanalytesvalidationofthelithiumheparinbdbarricorbloodcollectiontubeonahighlyautomatedrochecobas8000instrument AT locatellimassimo biochemicalimmunochemicalandserologyanalytesvalidationofthelithiumheparinbdbarricorbloodcollectiontubeonahighlyautomatedrochecobas8000instrument |