Biochemical, immunochemical and serology analytes validation of the lithium heparin BD Barricor blood collection tube on a highly automated Roche COBAS8000 instrument
BACKGROUND: Recently developed blood tubes with a barrier to provide plasma are becoming widespread. We compared 43 biochemical, 35 immunochemical and 7 serology analytes in a BD-Vacutainer® Barricor tube for local clinical validation of this lithium-heparin tube with a barrier. METHODS: Samples fro...
Autores principales: | Ferrari, Davide, Strollo, Marta, Vidali, Matteo, Motta, Andrea, Pontillo, Marina, Locatelli, Massimo |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
Mattioli 1885
2020
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC7569594/ https://www.ncbi.nlm.nih.gov/pubmed/32191654 http://dx.doi.org/10.23750/abm.v91i1.9195 |
Ejemplares similares
-
Evaluation of the Sentinel-FOB gold faecal immunochemical test for the presence of haemoglobin using the automated Roche Cobas 8000 system()
por: Schwettmann, Lutz, et al.
Publicado: (2022) -
The local clinical validation of a new lithium heparin tube with a barrier: BD Vacutainer® Barricor LH Plasma tube
por: Arslan, Fatma Demet, et al.
Publicado: (2017) -
Improved assay protocol for measurement of ammonia on the Roche Cobas 8000 automated platform
por: Kaplon, J., et al.
Publicado: (2018) -
Effect of specimen type on free immunoglobulin light chains analysis on the Roche Diagnostics cobas 8000 analyzer
por: Nelson, Louis S., et al.
Publicado: (2015) -
Test results comparison and sample stability study: is the BD Barricor tube a suitable replacement for the BD RST tube?
por: Mandić, Sanja, et al.
Publicado: (2020)