Cargando…
AAI to DDD pacing mode inappropriate switches in a dual chamber pacemaker due to ventricular blanking period
A 90-year-old woman received a dual chamber pacemaker (PM) for a sick sinus syndrome. The PM was programmed with SafeR AAI-DD pacing mode at 60 bpm. During a standard follow up, some memorized electrograms (EGMs) were found in SafeR diagnostics, with atrial pacing (Ap) not followed by any ventricula...
Autores principales: | , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
Elsevier
2020
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC7854375/ https://www.ncbi.nlm.nih.gov/pubmed/33221528 http://dx.doi.org/10.1016/j.ipej.2020.11.016 |