Heating of hip joint implants in MRI: The combined effect of RF and switched‐gradient fields

PURPOSE: To investigate how the simultaneous exposure to gradient and RF fields affects the temperature rise in patients with a metallic hip prosthesis during an MRI session. METHODS: In silico analysis was performed with an anatomically realistic human model with CoCrMo hip implant in 12 imaging po...