Cargando…

An In Vitro Biomechanical Evaluation of a Lateral Lumbar Interbody Fusion Device With Integrated Lateral Modular Plate Fixation

STUDY DESIGN: In vitro cadaveric biomechanical study. OBJECTIVE: Biomechanically characterize a novel lateral lumbar interbody fusion (LLIF) implant possessing integrated lateral modular plate fixation (MPF). METHODS: A human lumbar cadaveric (n = 7, L1-L4) biomechanical study of segmental range-of-...

Descripción completa

Detalles Bibliográficos
Autores principales: DenHaese, Ryan, Gandhi, Anup, Ferry, Chris, Farmer, Sam, Porter, Randall
Formato: Online Artículo Texto
Lenguaje:English
Publicado: SAGE Publications 2020
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC8013934/
https://www.ncbi.nlm.nih.gov/pubmed/32875868
http://dx.doi.org/10.1177/2192568220905611