Cargando…

A Power Tracking Algorithm for Early Detection of Centrifugal Flow Pump Thrombosis

Logfiles from the HeartWare HVAD System provide operational pump trend data to aid in patient management. Pump thrombosis is commonly associated with increases in the logfile power that may precede the clinical presentation. A Power Tracking algorithm was developed to detect significant deviations i...

Descripción completa

Detalles Bibliográficos
Autores principales: Slaughter, Mark S., Schlöglhofer, Thomas, Rich, Jonathan D., Brown, Michael C., Kadrolkar, Abhijit, Ramos, Veronica, Stadler, Robert W., Uriel, Nir, Mahr, Claudius, Sauer, Andrew J.
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Lippincott Williams & Wilkins 2021
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC8404954/
https://www.ncbi.nlm.nih.gov/pubmed/34225279