Cargando…

Biopsy Proven Focal Osteolysis in a Stainless-Steel Limb-Lengthening Device: A Report of Three Cases

Three pediatric patients presented with histologically confirmed osteolysis after limb lengthening with a magnetic, telescoping, stainless-steel device. The first patient's findings were discovered radiographically before routine removal of the device. In all cases, intraoperative histologic sp...

Descripción completa

Detalles Bibliográficos
Autores principales: Sax, Oliver C., Molavi, Diana W., Herzenberg, John E., Standard, Shawn C., McClure, Philip K.
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Wolters Kluwer 2021
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC8517308/
https://www.ncbi.nlm.nih.gov/pubmed/34644273
http://dx.doi.org/10.5435/JAAOSGlobal-D-21-00101