Biopsy Proven Focal Osteolysis in a Stainless-Steel Limb-Lengthening Device: A Report of Three Cases

Three pediatric patients presented with histologically confirmed osteolysis after limb lengthening with a magnetic, telescoping, stainless-steel device. The first patient's findings were discovered radiographically before routine removal of the device. In all cases, intraoperative histologic sp...

Descripción completa

Detalles Bibliográficos
Autores principales: Sax, Oliver C., Molavi, Diana W., Herzenberg, John E., Standard, Shawn C., McClure, Philip K.
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Wolters Kluwer 2021
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC8517308/
https://www.ncbi.nlm.nih.gov/pubmed/34644273
http://dx.doi.org/10.5435/JAAOSGlobal-D-21-00101
_version_ 1784583987477348352
author Sax, Oliver C.
Molavi, Diana W.
Herzenberg, John E.
Standard, Shawn C.
McClure, Philip K.
author_facet Sax, Oliver C.
Molavi, Diana W.
Herzenberg, John E.
Standard, Shawn C.
McClure, Philip K.
author_sort Sax, Oliver C.
collection PubMed
description Three pediatric patients presented with histologically confirmed osteolysis after limb lengthening with a magnetic, telescoping, stainless-steel device. The first patient's findings were discovered radiographically before routine removal of the device. In all cases, intraoperative histologic specimens taken from around the modular junction demonstrated particle-laden macrophages with suspicion for metal debris. Silicone debris was also identified. We found definitive osteolysis secondary to metal at the modular junction of three stainless-steel lengthening implants. This process is not well-understood in the setting of limb lengthening and should be examined further.
format Online
Article
Text
id pubmed-8517308
institution National Center for Biotechnology Information
language English
publishDate 2021
publisher Wolters Kluwer
record_format MEDLINE/PubMed
spelling pubmed-85173082021-10-18 Biopsy Proven Focal Osteolysis in a Stainless-Steel Limb-Lengthening Device: A Report of Three Cases Sax, Oliver C. Molavi, Diana W. Herzenberg, John E. Standard, Shawn C. McClure, Philip K. J Am Acad Orthop Surg Glob Res Rev Case Report Three pediatric patients presented with histologically confirmed osteolysis after limb lengthening with a magnetic, telescoping, stainless-steel device. The first patient's findings were discovered radiographically before routine removal of the device. In all cases, intraoperative histologic specimens taken from around the modular junction demonstrated particle-laden macrophages with suspicion for metal debris. Silicone debris was also identified. We found definitive osteolysis secondary to metal at the modular junction of three stainless-steel lengthening implants. This process is not well-understood in the setting of limb lengthening and should be examined further. Wolters Kluwer 2021-10-13 /pmc/articles/PMC8517308/ /pubmed/34644273 http://dx.doi.org/10.5435/JAAOSGlobal-D-21-00101 Text en Copyright © 2021 The Authors. Published by Wolters Kluwer Health, Inc. on behalf of the American Academy of Orthopaedic Surgeons. https://creativecommons.org/licenses/by/4.0/This is an open access article distributed under the Creative Commons Attribution License 4.0 (CCBY) (https://creativecommons.org/licenses/by/4.0/) , which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited.
spellingShingle Case Report
Sax, Oliver C.
Molavi, Diana W.
Herzenberg, John E.
Standard, Shawn C.
McClure, Philip K.
Biopsy Proven Focal Osteolysis in a Stainless-Steel Limb-Lengthening Device: A Report of Three Cases
title Biopsy Proven Focal Osteolysis in a Stainless-Steel Limb-Lengthening Device: A Report of Three Cases
title_full Biopsy Proven Focal Osteolysis in a Stainless-Steel Limb-Lengthening Device: A Report of Three Cases
title_fullStr Biopsy Proven Focal Osteolysis in a Stainless-Steel Limb-Lengthening Device: A Report of Three Cases
title_full_unstemmed Biopsy Proven Focal Osteolysis in a Stainless-Steel Limb-Lengthening Device: A Report of Three Cases
title_short Biopsy Proven Focal Osteolysis in a Stainless-Steel Limb-Lengthening Device: A Report of Three Cases
title_sort biopsy proven focal osteolysis in a stainless-steel limb-lengthening device: a report of three cases
topic Case Report
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC8517308/
https://www.ncbi.nlm.nih.gov/pubmed/34644273
http://dx.doi.org/10.5435/JAAOSGlobal-D-21-00101
work_keys_str_mv AT saxoliverc biopsyprovenfocalosteolysisinastainlesssteellimblengtheningdeviceareportofthreecases
AT molavidianaw biopsyprovenfocalosteolysisinastainlesssteellimblengtheningdeviceareportofthreecases
AT herzenbergjohne biopsyprovenfocalosteolysisinastainlesssteellimblengtheningdeviceareportofthreecases
AT standardshawnc biopsyprovenfocalosteolysisinastainlesssteellimblengtheningdeviceareportofthreecases
AT mcclurephilipk biopsyprovenfocalosteolysisinastainlesssteellimblengtheningdeviceareportofthreecases