Smartphone application-supported validation of three automatic devices for self-measurement of blood pressure according to the European Society of Hypertension International Protocol revision 2010: the Omron HEM-7120, Yuwell YE680A and Cofoe KF-65B
BACKGROUND: Accurate measurement of blood pressure (BP) is crucial to hypertension control and prevention of future stroke and heart attack. All BP measuring devices must be validated independently in the clinical setting. OBJECTIVE: To validate the accuracy of three automatic upper arm devices (Omr...