In-House Algorithm for Reporting Discrepant HbA1c Result and Troubleshooting a Case of False Low HbA1c

We report an unusual case of a patient having low glycosylated hemoglobin (HbA1c) below the reportable range, despite having borderline fasting blood glucose. The patient had decreased erythrocytes count and elevated reticulocyte count, with no evidence of hemoglobinopathy. He reported taking multid...

Descripción completa

Detalles Bibliográficos
Autores principales: Pant, Vivek, Shrestha, Abha, Pyakurel, Devish, Gautam, Keyoor, Pradhan, Santosh
Formato: Online Artículo Texto
Lenguaje:English
Publicado: The Communications and Publications Division (CPD) of the IFCC 2021
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC8592636/
https://www.ncbi.nlm.nih.gov/pubmed/34819826
_version_ 1784599506874007552
author Pant, Vivek
Shrestha, Abha
Pyakurel, Devish
Gautam, Keyoor
Pradhan, Santosh
author_facet Pant, Vivek
Shrestha, Abha
Pyakurel, Devish
Gautam, Keyoor
Pradhan, Santosh
author_sort Pant, Vivek
collection PubMed
description We report an unusual case of a patient having low glycosylated hemoglobin (HbA1c) below the reportable range, despite having borderline fasting blood glucose. The patient had decreased erythrocytes count and elevated reticulocyte count, with no evidence of hemoglobinopathy. He reported taking multidrug therapy for borderline lepromatous leprosy. Dapsone induced hemolysis was identified as the cause for the discordant HbA1c. Thus, it is important to be aware of medications and conditions that may lead to a falsely low HbA1c level so that incorrect treatment decisions are not made. In such situations, alternative measure of glycemic control, such as fructosamine is recommended. Further it is also recommended that clinical laboratories have standard protocol to troubleshoot any discrepant HbA1c result.
format Online
Article
Text
id pubmed-8592636
institution National Center for Biotechnology Information
language English
publishDate 2021
publisher The Communications and Publications Division (CPD) of the IFCC
record_format MEDLINE/PubMed
spelling pubmed-85926362021-11-23 In-House Algorithm for Reporting Discrepant HbA1c Result and Troubleshooting a Case of False Low HbA1c Pant, Vivek Shrestha, Abha Pyakurel, Devish Gautam, Keyoor Pradhan, Santosh EJIFCC Research Article We report an unusual case of a patient having low glycosylated hemoglobin (HbA1c) below the reportable range, despite having borderline fasting blood glucose. The patient had decreased erythrocytes count and elevated reticulocyte count, with no evidence of hemoglobinopathy. He reported taking multidrug therapy for borderline lepromatous leprosy. Dapsone induced hemolysis was identified as the cause for the discordant HbA1c. Thus, it is important to be aware of medications and conditions that may lead to a falsely low HbA1c level so that incorrect treatment decisions are not made. In such situations, alternative measure of glycemic control, such as fructosamine is recommended. Further it is also recommended that clinical laboratories have standard protocol to troubleshoot any discrepant HbA1c result. The Communications and Publications Division (CPD) of the IFCC 2021-10-18 /pmc/articles/PMC8592636/ /pubmed/34819826 Text en Copyright © 2021 International Federation of Clinical Chemistry and Laboratory Medicine (IFCC). All rights reserved. https://creativecommons.org/licenses/by-nc/4.0/This is a Platinum Open Access Journal distributed under the terms of the Creative Commons Attribution Non-Commercial License which permits unrestricted non-commercial use, distribution, and reproduction in any medium, provided the original work is properly cited.
spellingShingle Research Article
Pant, Vivek
Shrestha, Abha
Pyakurel, Devish
Gautam, Keyoor
Pradhan, Santosh
In-House Algorithm for Reporting Discrepant HbA1c Result and Troubleshooting a Case of False Low HbA1c
title In-House Algorithm for Reporting Discrepant HbA1c Result and Troubleshooting a Case of False Low HbA1c
title_full In-House Algorithm for Reporting Discrepant HbA1c Result and Troubleshooting a Case of False Low HbA1c
title_fullStr In-House Algorithm for Reporting Discrepant HbA1c Result and Troubleshooting a Case of False Low HbA1c
title_full_unstemmed In-House Algorithm for Reporting Discrepant HbA1c Result and Troubleshooting a Case of False Low HbA1c
title_short In-House Algorithm for Reporting Discrepant HbA1c Result and Troubleshooting a Case of False Low HbA1c
title_sort in-house algorithm for reporting discrepant hba1c result and troubleshooting a case of false low hba1c
topic Research Article
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC8592636/
https://www.ncbi.nlm.nih.gov/pubmed/34819826
work_keys_str_mv AT pantvivek inhousealgorithmforreportingdiscrepanthba1cresultandtroubleshootingacaseoffalselowhba1c
AT shresthaabha inhousealgorithmforreportingdiscrepanthba1cresultandtroubleshootingacaseoffalselowhba1c
AT pyakureldevish inhousealgorithmforreportingdiscrepanthba1cresultandtroubleshootingacaseoffalselowhba1c
AT gautamkeyoor inhousealgorithmforreportingdiscrepanthba1cresultandtroubleshootingacaseoffalselowhba1c
AT pradhansantosh inhousealgorithmforreportingdiscrepanthba1cresultandtroubleshootingacaseoffalselowhba1c