Cargando…
Evaluation of the Ultrastructural and In Vitro Flow Properties of the PRESERFLO MicroShunt
PURPOSE: To measure the in vitro flow properties of the PRESERFLO implant for comparison with the theoretical resistance to flow. METHODS: The PRESERFLO was designed to control the flow of aqueous humor according to the Hagen-Poiseuille (HP) equation. Scanning electron microscopy (SEM) was performed...
Autores principales: | , , , , , , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
The Association for Research in Vision and Ophthalmology
2021
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC8606849/ https://www.ncbi.nlm.nih.gov/pubmed/34792556 http://dx.doi.org/10.1167/tvst.10.13.26 |