Cargando…

Chromatography at −30 °C for Reduced Back-Exchange, Reduced Carryover, and Improved Dynamic Range for Hydrogen–Deuterium Exchange Mass Spectrometry

[Image: see text] For hydrogen–deuterium exchange mass spectrometry (HDX-MS) to have an increased role in quality control of biopharmaceuticals, H for D back-exchange occurring during protein analyses should be minimized to promote greater reproducibility. Standard HDX-MS analysis systems that diges...

Descripción completa

Detalles Bibliográficos
Autores principales: Anderson, Kyle W., Hudgens, Jeffrey W.
Formato: Online Artículo Texto
Lenguaje:English
Publicado: American Chemical Society 2022
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC9264389/
https://www.ncbi.nlm.nih.gov/pubmed/35732031
http://dx.doi.org/10.1021/jasms.2c00096
_version_ 1784742966903963648
author Anderson, Kyle W.
Hudgens, Jeffrey W.
author_facet Anderson, Kyle W.
Hudgens, Jeffrey W.
author_sort Anderson, Kyle W.
collection PubMed
description [Image: see text] For hydrogen–deuterium exchange mass spectrometry (HDX-MS) to have an increased role in quality control of biopharmaceuticals, H for D back-exchange occurring during protein analyses should be minimized to promote greater reproducibility. Standard HDX-MS analysis systems that digest proteins and separate peptides at pH 2.7 and 0 °C can lose >30% of the deuterium marker within 15 min of sample injection. This report describes the architecture and performance of a dual-enzyme, HDX-MS instrument that conducts liquid chromatography (LC) separations at subzero temperature, thereby reducing back-exchange and supporting longer LC separations with improved chromatographic resolution. LC separations of perdeuterated, fully reduced, iodoacetamide-treated BSA protein digest standard peptides were performed at 0, −10, −20, and −30 °C in ethylene glycol (EG)/H(2)O mixtures. Analyses conducted at −20 and −30 °C produced similar results. After subtracting for deuterium retained in arginine side chains, the average peptide eluted during a 40 min gradient contained ≈16% more deuterium than peptides eluted with a conventional 8 min gradient at 0 °C. A subset of peptides exhibited ≈26% more deuterium. Although chromatographic peaks shift with EG concentration and temperature, the apparatus elutes unbroadened LC peaks. Electrospray ion intensity does not decline with increasing EG fraction. To minimize bias from sample carryover, the fluidic circuits allow flush and backflush cleaning of all enzyme and LC columns. The system can perform LC separations and clean enzyme columns simultaneously. Temperature zones are controlled ±0.058 °C. The potential of increased sensitivity by mixing acetonitrile with the analytical column effluent was also examined.
format Online
Article
Text
id pubmed-9264389
institution National Center for Biotechnology Information
language English
publishDate 2022
publisher American Chemical Society
record_format MEDLINE/PubMed
spelling pubmed-92643892022-07-09 Chromatography at −30 °C for Reduced Back-Exchange, Reduced Carryover, and Improved Dynamic Range for Hydrogen–Deuterium Exchange Mass Spectrometry Anderson, Kyle W. Hudgens, Jeffrey W. J Am Soc Mass Spectrom [Image: see text] For hydrogen–deuterium exchange mass spectrometry (HDX-MS) to have an increased role in quality control of biopharmaceuticals, H for D back-exchange occurring during protein analyses should be minimized to promote greater reproducibility. Standard HDX-MS analysis systems that digest proteins and separate peptides at pH 2.7 and 0 °C can lose >30% of the deuterium marker within 15 min of sample injection. This report describes the architecture and performance of a dual-enzyme, HDX-MS instrument that conducts liquid chromatography (LC) separations at subzero temperature, thereby reducing back-exchange and supporting longer LC separations with improved chromatographic resolution. LC separations of perdeuterated, fully reduced, iodoacetamide-treated BSA protein digest standard peptides were performed at 0, −10, −20, and −30 °C in ethylene glycol (EG)/H(2)O mixtures. Analyses conducted at −20 and −30 °C produced similar results. After subtracting for deuterium retained in arginine side chains, the average peptide eluted during a 40 min gradient contained ≈16% more deuterium than peptides eluted with a conventional 8 min gradient at 0 °C. A subset of peptides exhibited ≈26% more deuterium. Although chromatographic peaks shift with EG concentration and temperature, the apparatus elutes unbroadened LC peaks. Electrospray ion intensity does not decline with increasing EG fraction. To minimize bias from sample carryover, the fluidic circuits allow flush and backflush cleaning of all enzyme and LC columns. The system can perform LC separations and clean enzyme columns simultaneously. Temperature zones are controlled ±0.058 °C. The potential of increased sensitivity by mixing acetonitrile with the analytical column effluent was also examined. American Chemical Society 2022-06-22 2022-07-06 /pmc/articles/PMC9264389/ /pubmed/35732031 http://dx.doi.org/10.1021/jasms.2c00096 Text en Not subject to U.S. Copyright. Published 2022 by American Chemical Society https://creativecommons.org/licenses/by-nc-nd/4.0/Permits non-commercial access and re-use, provided that author attribution and integrity are maintained; but does not permit creation of adaptations or other derivative works (https://creativecommons.org/licenses/by-nc-nd/4.0/).
spellingShingle Anderson, Kyle W.
Hudgens, Jeffrey W.
Chromatography at −30 °C for Reduced Back-Exchange, Reduced Carryover, and Improved Dynamic Range for Hydrogen–Deuterium Exchange Mass Spectrometry
title Chromatography at −30 °C for Reduced Back-Exchange, Reduced Carryover, and Improved Dynamic Range for Hydrogen–Deuterium Exchange Mass Spectrometry
title_full Chromatography at −30 °C for Reduced Back-Exchange, Reduced Carryover, and Improved Dynamic Range for Hydrogen–Deuterium Exchange Mass Spectrometry
title_fullStr Chromatography at −30 °C for Reduced Back-Exchange, Reduced Carryover, and Improved Dynamic Range for Hydrogen–Deuterium Exchange Mass Spectrometry
title_full_unstemmed Chromatography at −30 °C for Reduced Back-Exchange, Reduced Carryover, and Improved Dynamic Range for Hydrogen–Deuterium Exchange Mass Spectrometry
title_short Chromatography at −30 °C for Reduced Back-Exchange, Reduced Carryover, and Improved Dynamic Range for Hydrogen–Deuterium Exchange Mass Spectrometry
title_sort chromatography at −30 °c for reduced back-exchange, reduced carryover, and improved dynamic range for hydrogen–deuterium exchange mass spectrometry
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC9264389/
https://www.ncbi.nlm.nih.gov/pubmed/35732031
http://dx.doi.org/10.1021/jasms.2c00096
work_keys_str_mv AT andersonkylew chromatographyat30cforreducedbackexchangereducedcarryoverandimproveddynamicrangeforhydrogendeuteriumexchangemassspectrometry
AT hudgensjeffreyw chromatographyat30cforreducedbackexchangereducedcarryoverandimproveddynamicrangeforhydrogendeuteriumexchangemassspectrometry