Cargando…
Cannula Fracture during Transoral Endoscopic Thyroidectomy Vestibular Approach: Causes and Prevention
In the transoral endoscopic thyroidectomy vestibular approach (TOETVA), three oral vestibular incisions are used to access the thyroid. This approach leaves no scar on the body surface; however, unexpected complications may occur. Three patients (two women, one man) underwent TOETVA using the standa...
Autores principales: | , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
MDPI
2022
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC9320644/ https://www.ncbi.nlm.nih.gov/pubmed/35885472 http://dx.doi.org/10.3390/diagnostics12071566 |
_version_ | 1784755842249129984 |
---|---|
author | Liang, Tsung-Jung Liu, Shiuh-Inn Chen, I-Shu |
author_facet | Liang, Tsung-Jung Liu, Shiuh-Inn Chen, I-Shu |
author_sort | Liang, Tsung-Jung |
collection | PubMed |
description | In the transoral endoscopic thyroidectomy vestibular approach (TOETVA), three oral vestibular incisions are used to access the thyroid. This approach leaves no scar on the body surface; however, unexpected complications may occur. Three patients (two women, one man) underwent TOETVA using the standard three-port technique. Broken cannulas of the 12 mm central port were noted in all cases. All cannulas broke on the ventral side of the distal shaft. The fracture lines were 3–4 cm in length, with some fragments scattered throughout the operative field and oral cavity. The fractures were caused by compression against the mandible while tilting the cannula during surgical manipulation. Male sex, short stature, and protruding chin may be risk factors for cannula fracture in TOETVA. Measures should be taken to prevent this complication, particularly in high-risk patients. |
format | Online Article Text |
id | pubmed-9320644 |
institution | National Center for Biotechnology Information |
language | English |
publishDate | 2022 |
publisher | MDPI |
record_format | MEDLINE/PubMed |
spelling | pubmed-93206442022-07-27 Cannula Fracture during Transoral Endoscopic Thyroidectomy Vestibular Approach: Causes and Prevention Liang, Tsung-Jung Liu, Shiuh-Inn Chen, I-Shu Diagnostics (Basel) Case Report In the transoral endoscopic thyroidectomy vestibular approach (TOETVA), three oral vestibular incisions are used to access the thyroid. This approach leaves no scar on the body surface; however, unexpected complications may occur. Three patients (two women, one man) underwent TOETVA using the standard three-port technique. Broken cannulas of the 12 mm central port were noted in all cases. All cannulas broke on the ventral side of the distal shaft. The fracture lines were 3–4 cm in length, with some fragments scattered throughout the operative field and oral cavity. The fractures were caused by compression against the mandible while tilting the cannula during surgical manipulation. Male sex, short stature, and protruding chin may be risk factors for cannula fracture in TOETVA. Measures should be taken to prevent this complication, particularly in high-risk patients. MDPI 2022-06-28 /pmc/articles/PMC9320644/ /pubmed/35885472 http://dx.doi.org/10.3390/diagnostics12071566 Text en © 2022 by the authors. https://creativecommons.org/licenses/by/4.0/Licensee MDPI, Basel, Switzerland. This article is an open access article distributed under the terms and conditions of the Creative Commons Attribution (CC BY) license (https://creativecommons.org/licenses/by/4.0/). |
spellingShingle | Case Report Liang, Tsung-Jung Liu, Shiuh-Inn Chen, I-Shu Cannula Fracture during Transoral Endoscopic Thyroidectomy Vestibular Approach: Causes and Prevention |
title | Cannula Fracture during Transoral Endoscopic Thyroidectomy Vestibular Approach: Causes and Prevention |
title_full | Cannula Fracture during Transoral Endoscopic Thyroidectomy Vestibular Approach: Causes and Prevention |
title_fullStr | Cannula Fracture during Transoral Endoscopic Thyroidectomy Vestibular Approach: Causes and Prevention |
title_full_unstemmed | Cannula Fracture during Transoral Endoscopic Thyroidectomy Vestibular Approach: Causes and Prevention |
title_short | Cannula Fracture during Transoral Endoscopic Thyroidectomy Vestibular Approach: Causes and Prevention |
title_sort | cannula fracture during transoral endoscopic thyroidectomy vestibular approach: causes and prevention |
topic | Case Report |
url | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC9320644/ https://www.ncbi.nlm.nih.gov/pubmed/35885472 http://dx.doi.org/10.3390/diagnostics12071566 |
work_keys_str_mv | AT liangtsungjung cannulafractureduringtransoralendoscopicthyroidectomyvestibularapproachcausesandprevention AT liushiuhinn cannulafractureduringtransoralendoscopicthyroidectomyvestibularapproachcausesandprevention AT chenishu cannulafractureduringtransoralendoscopicthyroidectomyvestibularapproachcausesandprevention |