Cargando…

Cannula Fracture during Transoral Endoscopic Thyroidectomy Vestibular Approach: Causes and Prevention

In the transoral endoscopic thyroidectomy vestibular approach (TOETVA), three oral vestibular incisions are used to access the thyroid. This approach leaves no scar on the body surface; however, unexpected complications may occur. Three patients (two women, one man) underwent TOETVA using the standa...

Descripción completa

Detalles Bibliográficos
Autores principales: Liang, Tsung-Jung, Liu, Shiuh-Inn, Chen, I-Shu
Formato: Online Artículo Texto
Lenguaje:English
Publicado: MDPI 2022
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC9320644/
https://www.ncbi.nlm.nih.gov/pubmed/35885472
http://dx.doi.org/10.3390/diagnostics12071566
_version_ 1784755842249129984
author Liang, Tsung-Jung
Liu, Shiuh-Inn
Chen, I-Shu
author_facet Liang, Tsung-Jung
Liu, Shiuh-Inn
Chen, I-Shu
author_sort Liang, Tsung-Jung
collection PubMed
description In the transoral endoscopic thyroidectomy vestibular approach (TOETVA), three oral vestibular incisions are used to access the thyroid. This approach leaves no scar on the body surface; however, unexpected complications may occur. Three patients (two women, one man) underwent TOETVA using the standard three-port technique. Broken cannulas of the 12 mm central port were noted in all cases. All cannulas broke on the ventral side of the distal shaft. The fracture lines were 3–4 cm in length, with some fragments scattered throughout the operative field and oral cavity. The fractures were caused by compression against the mandible while tilting the cannula during surgical manipulation. Male sex, short stature, and protruding chin may be risk factors for cannula fracture in TOETVA. Measures should be taken to prevent this complication, particularly in high-risk patients.
format Online
Article
Text
id pubmed-9320644
institution National Center for Biotechnology Information
language English
publishDate 2022
publisher MDPI
record_format MEDLINE/PubMed
spelling pubmed-93206442022-07-27 Cannula Fracture during Transoral Endoscopic Thyroidectomy Vestibular Approach: Causes and Prevention Liang, Tsung-Jung Liu, Shiuh-Inn Chen, I-Shu Diagnostics (Basel) Case Report In the transoral endoscopic thyroidectomy vestibular approach (TOETVA), three oral vestibular incisions are used to access the thyroid. This approach leaves no scar on the body surface; however, unexpected complications may occur. Three patients (two women, one man) underwent TOETVA using the standard three-port technique. Broken cannulas of the 12 mm central port were noted in all cases. All cannulas broke on the ventral side of the distal shaft. The fracture lines were 3–4 cm in length, with some fragments scattered throughout the operative field and oral cavity. The fractures were caused by compression against the mandible while tilting the cannula during surgical manipulation. Male sex, short stature, and protruding chin may be risk factors for cannula fracture in TOETVA. Measures should be taken to prevent this complication, particularly in high-risk patients. MDPI 2022-06-28 /pmc/articles/PMC9320644/ /pubmed/35885472 http://dx.doi.org/10.3390/diagnostics12071566 Text en © 2022 by the authors. https://creativecommons.org/licenses/by/4.0/Licensee MDPI, Basel, Switzerland. This article is an open access article distributed under the terms and conditions of the Creative Commons Attribution (CC BY) license (https://creativecommons.org/licenses/by/4.0/).
spellingShingle Case Report
Liang, Tsung-Jung
Liu, Shiuh-Inn
Chen, I-Shu
Cannula Fracture during Transoral Endoscopic Thyroidectomy Vestibular Approach: Causes and Prevention
title Cannula Fracture during Transoral Endoscopic Thyroidectomy Vestibular Approach: Causes and Prevention
title_full Cannula Fracture during Transoral Endoscopic Thyroidectomy Vestibular Approach: Causes and Prevention
title_fullStr Cannula Fracture during Transoral Endoscopic Thyroidectomy Vestibular Approach: Causes and Prevention
title_full_unstemmed Cannula Fracture during Transoral Endoscopic Thyroidectomy Vestibular Approach: Causes and Prevention
title_short Cannula Fracture during Transoral Endoscopic Thyroidectomy Vestibular Approach: Causes and Prevention
title_sort cannula fracture during transoral endoscopic thyroidectomy vestibular approach: causes and prevention
topic Case Report
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC9320644/
https://www.ncbi.nlm.nih.gov/pubmed/35885472
http://dx.doi.org/10.3390/diagnostics12071566
work_keys_str_mv AT liangtsungjung cannulafractureduringtransoralendoscopicthyroidectomyvestibularapproachcausesandprevention
AT liushiuhinn cannulafractureduringtransoralendoscopicthyroidectomyvestibularapproachcausesandprevention
AT chenishu cannulafractureduringtransoralendoscopicthyroidectomyvestibularapproachcausesandprevention