Cargando…

Improving PMUT Receive Sensitivity via DC Bias and Piezoelectric Composition

The receive sensitivity of lead zirconate titanate (PZT) piezoelectric micromachined ultrasound transducers (PMUTs) was improved by applying a DC bias during operation. The PMUT receive sensitivity is governed by the voltage piezoelectric coefficient, h(31,f). With applied DC biases (up to 15 V) on...

Descripción completa

Detalles Bibliográficos
Autores principales: Cheng, Christopher, Peters, Travis, Dangi, Ajay, Agrawal, Sumit, Chen, Haoyang, Kothapalli, Sri-Rajasekhar, Trolier-McKinstry, Susan
Formato: Online Artículo Texto
Lenguaje:English
Publicado: MDPI 2022
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC9370952/
https://www.ncbi.nlm.nih.gov/pubmed/35957175
http://dx.doi.org/10.3390/s22155614
Descripción
Sumario:The receive sensitivity of lead zirconate titanate (PZT) piezoelectric micromachined ultrasound transducers (PMUTs) was improved by applying a DC bias during operation. The PMUT receive sensitivity is governed by the voltage piezoelectric coefficient, h(31,f). With applied DC biases (up to 15 V) on a 2 μm PbZr(0.52)Ti(0.48)O(3) film, e(31,f) increased 1.6 times, permittivity decreased by a factor of 0.6, and the voltage coefficient increased by ~2.5 times. For released PMUT devices, the ultrasound receive sensitivity improved by 2.5 times and the photoacoustic signal improved 1.9 times with 15 V applied DC bias. B-mode photoacoustic imaging experiments showed that with DC bias, the PMUT received clearer photoacoustic signals from pencil leads at 4.3 cm, compared to 3.7 cm without DC bias.