Cargando…

Long-term durability of a new surgical aortic valve: A 1 billion cycle in vitro study

BACKGROUND: This study assessed the long-term hemodynamic functional performance of the new Inspiris Resilia aortic valve after accelerated wear testing (AWT). METHODS: Three 21-mm and 23-mm Inspiris valves were used for the AWT procedure. After 1 billion cycles (equivalent to 25 years), the valves&...

Descripción completa

Detalles Bibliográficos
Autores principales: Sadri, Vahid, Trusty, Phillip M., Madukauwa-David, Immanuel David, Yoganathan, Ajit P.
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Elsevier 2021
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC9390709/
https://www.ncbi.nlm.nih.gov/pubmed/36003435
http://dx.doi.org/10.1016/j.xjon.2021.10.056