Long-term durability of a new surgical aortic valve: A 1 billion cycle in vitro study
BACKGROUND: This study assessed the long-term hemodynamic functional performance of the new Inspiris Resilia aortic valve after accelerated wear testing (AWT). METHODS: Three 21-mm and 23-mm Inspiris valves were used for the AWT procedure. After 1 billion cycles (equivalent to 25 years), the valves&...
Autores principales: | , , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
Elsevier
2021
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC9390709/ https://www.ncbi.nlm.nih.gov/pubmed/36003435 http://dx.doi.org/10.1016/j.xjon.2021.10.056 |
_version_ | 1784770714464681984 |
---|---|
author | Sadri, Vahid Trusty, Phillip M. Madukauwa-David, Immanuel David Yoganathan, Ajit P. |
author_facet | Sadri, Vahid Trusty, Phillip M. Madukauwa-David, Immanuel David Yoganathan, Ajit P. |
author_sort | Sadri, Vahid |
collection | PubMed |
description | BACKGROUND: This study assessed the long-term hemodynamic functional performance of the new Inspiris Resilia aortic valve after accelerated wear testing (AWT). METHODS: Three 21-mm and 23-mm Inspiris valves were used for the AWT procedure. After 1 billion cycles (equivalent to 25 years), the valves' hemodynamic performance was compared with that of the corresponding zero-cycled condition. Next, 1 AWT cycled valve of each valve size was selected at random for particle image velocimetry (PIV) and leaflet kinematic tests, and the data were compared with data for an uncycled Inspiris Resilia aortic valve of the same size. PIV was used to quantitatively evaluate flow fields downstream of the valve. Valves were tested according to International Standards Organization 5840-2:2015 protocols. RESULTS: The 21-mm and 23-mm valves met the International Organization for Standardization (ISO) durability performance requirements to 1 billion cycles. The mean effective orifice areas for the 21-mm and 23-mm zero-cycled and 1 billion–cycled valves were 1.89 ± 0.02 cm(2) and 1.94 ± 0.01 cm(2), respectively (P < .05) and 2.3 ± 0.13 cm(2) and 2.40 ± 0.11 cm(2), respectively (P < .05). Flow characterization of the control valves and the study valves demonstrated similar flow characteristics. The velocity and shear stress fields were also similar in the control and study valves. CONCLUSIONS: The Inspiris Resilia aortic valve demonstrated very good durability and hemodynamic performance after an equivalent of 25 years of simulated in vitro accelerated wear. The study valves exceeded 1 billion cycles of simulated wear, 5 times longer than the standard requirement for a tissue valve as stipulated in ISO 5840-2:2015. |
format | Online Article Text |
id | pubmed-9390709 |
institution | National Center for Biotechnology Information |
language | English |
publishDate | 2021 |
publisher | Elsevier |
record_format | MEDLINE/PubMed |
spelling | pubmed-93907092022-08-23 Long-term durability of a new surgical aortic valve: A 1 billion cycle in vitro study Sadri, Vahid Trusty, Phillip M. Madukauwa-David, Immanuel David Yoganathan, Ajit P. JTCVS Open Adult: Aortic Valve BACKGROUND: This study assessed the long-term hemodynamic functional performance of the new Inspiris Resilia aortic valve after accelerated wear testing (AWT). METHODS: Three 21-mm and 23-mm Inspiris valves were used for the AWT procedure. After 1 billion cycles (equivalent to 25 years), the valves' hemodynamic performance was compared with that of the corresponding zero-cycled condition. Next, 1 AWT cycled valve of each valve size was selected at random for particle image velocimetry (PIV) and leaflet kinematic tests, and the data were compared with data for an uncycled Inspiris Resilia aortic valve of the same size. PIV was used to quantitatively evaluate flow fields downstream of the valve. Valves were tested according to International Standards Organization 5840-2:2015 protocols. RESULTS: The 21-mm and 23-mm valves met the International Organization for Standardization (ISO) durability performance requirements to 1 billion cycles. The mean effective orifice areas for the 21-mm and 23-mm zero-cycled and 1 billion–cycled valves were 1.89 ± 0.02 cm(2) and 1.94 ± 0.01 cm(2), respectively (P < .05) and 2.3 ± 0.13 cm(2) and 2.40 ± 0.11 cm(2), respectively (P < .05). Flow characterization of the control valves and the study valves demonstrated similar flow characteristics. The velocity and shear stress fields were also similar in the control and study valves. CONCLUSIONS: The Inspiris Resilia aortic valve demonstrated very good durability and hemodynamic performance after an equivalent of 25 years of simulated in vitro accelerated wear. The study valves exceeded 1 billion cycles of simulated wear, 5 times longer than the standard requirement for a tissue valve as stipulated in ISO 5840-2:2015. Elsevier 2021-11-04 /pmc/articles/PMC9390709/ /pubmed/36003435 http://dx.doi.org/10.1016/j.xjon.2021.10.056 Text en © 2021 The Author(s) https://creativecommons.org/licenses/by-nc-nd/4.0/This is an open access article under the CC BY-NC-ND license (http://creativecommons.org/licenses/by-nc-nd/4.0/). |
spellingShingle | Adult: Aortic Valve Sadri, Vahid Trusty, Phillip M. Madukauwa-David, Immanuel David Yoganathan, Ajit P. Long-term durability of a new surgical aortic valve: A 1 billion cycle in vitro study |
title | Long-term durability of a new surgical aortic valve: A 1 billion cycle in vitro study |
title_full | Long-term durability of a new surgical aortic valve: A 1 billion cycle in vitro study |
title_fullStr | Long-term durability of a new surgical aortic valve: A 1 billion cycle in vitro study |
title_full_unstemmed | Long-term durability of a new surgical aortic valve: A 1 billion cycle in vitro study |
title_short | Long-term durability of a new surgical aortic valve: A 1 billion cycle in vitro study |
title_sort | long-term durability of a new surgical aortic valve: a 1 billion cycle in vitro study |
topic | Adult: Aortic Valve |
url | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC9390709/ https://www.ncbi.nlm.nih.gov/pubmed/36003435 http://dx.doi.org/10.1016/j.xjon.2021.10.056 |
work_keys_str_mv | AT sadrivahid longtermdurabilityofanewsurgicalaorticvalvea1billioncycleinvitrostudy AT trustyphillipm longtermdurabilityofanewsurgicalaorticvalvea1billioncycleinvitrostudy AT madukauwadavidimmanueldavid longtermdurabilityofanewsurgicalaorticvalvea1billioncycleinvitrostudy AT yoganathanajitp longtermdurabilityofanewsurgicalaorticvalvea1billioncycleinvitrostudy |