Cargando…
Whole-blood Point-of-Care Activated Partial Thromboplastin Time Ratio (APR) is not Accurate Enough to Monitor Heparin Therapy in Patients with Severe Respiratory Failure Secondary to SARS-Cov-2 Infection Supported with Veno-Venous Extracorporeal Membrane Oxygenation (VV-ECMO)
Support with VV-ECMO requires anticoagulation with unfractionated heparin to prevent thrombotic complications. This must be monitored due to bleeding risk. A point-of-care (POC) method of testing aPTT and APR was evaluated for agreement with laboratory methods. In a prospective observational study,...