Cargando…
Small Footprint Biocompatible Antenna for Implantable Devices: Design, In-Silico, In-Vitro and Ex-Vivo Testing
In this paper, a miniaturized skin implantable biocompatible microstrip antenna is proposed for biotelemetry applications at 2.4–2.48 GHz. The volume of the antenna is 42.68 mm(3) with dimensions 8.2 × 6.94 × 0.75 mm(3). Rogers RO Duroid 3010 of 0.25 mm thickness is used for designing the proposed b...
Autores principales: | , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
Springer International Publishing
2023
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC9838436/ http://dx.doi.org/10.1007/s40998-023-00588-8 |