Overestimation of Oxygen Saturation Measured by Pulse Oximetry in Hypoxemia. Part 1: Effect of Optical Pathlengths-Ratio Increase

On average, arterial oxygen saturation measured by pulse oximetry (SpO(2)) is higher in hypoxemia than the true oxygen saturation measured invasively (SaO(2)), thereby increasing the risk of occult hypoxemia. In the current article, measurements of SpO(2) on 17 cyanotic newborns were performed by me...

Descripción completa

Detalles Bibliográficos
Autores principales: Elron, Eyal, Bromiker, Ruben, Gleisner, Ori, Yosef-Hai, Ohad, Goldberg, Ori, Nitzan, Itamar, Nitzan, Meir
Formato: Online Artículo Texto
Lenguaje:English
Publicado: MDPI 2023
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC9921559/
https://www.ncbi.nlm.nih.gov/pubmed/36772474
http://dx.doi.org/10.3390/s23031434