Overestimation of Oxygen Saturation Measured by Pulse Oximetry in Hypoxemia. Part 1: Effect of Optical Pathlengths-Ratio Increase

On average, arterial oxygen saturation measured by pulse oximetry (SpO(2)) is higher in hypoxemia than the true oxygen saturation measured invasively (SaO(2)), thereby increasing the risk of occult hypoxemia. In the current article, measurements of SpO(2) on 17 cyanotic newborns were performed by me...

Descripción completa

Detalles Bibliográficos
Autores principales: Elron, Eyal, Bromiker, Ruben, Gleisner, Ori, Yosef-Hai, Ohad, Goldberg, Ori, Nitzan, Itamar, Nitzan, Meir
Formato: Online Artículo Texto
Lenguaje:English
Publicado: MDPI 2023
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC9921559/
https://www.ncbi.nlm.nih.gov/pubmed/36772474
http://dx.doi.org/10.3390/s23031434
_version_ 1784887341090865152
author Elron, Eyal
Bromiker, Ruben
Gleisner, Ori
Yosef-Hai, Ohad
Goldberg, Ori
Nitzan, Itamar
Nitzan, Meir
author_facet Elron, Eyal
Bromiker, Ruben
Gleisner, Ori
Yosef-Hai, Ohad
Goldberg, Ori
Nitzan, Itamar
Nitzan, Meir
author_sort Elron, Eyal
collection PubMed
description On average, arterial oxygen saturation measured by pulse oximetry (SpO(2)) is higher in hypoxemia than the true oxygen saturation measured invasively (SaO(2)), thereby increasing the risk of occult hypoxemia. In the current article, measurements of SpO(2) on 17 cyanotic newborns were performed by means of a Nellcor pulse oximeter (POx), based on light with two wavelengths in the red and infrared regions (660 and 900 nm), and by means of a novel POx, based on two wavelengths in the infrared region (761 and 820 nm). The SpO(2) readings from the two POxs showed higher values than the invasive SaO(2) readings, and the disparity increased with decreasing SaO(2). SpO(2) measured using the two infrared wavelengths showed better correlation with SaO(2) than SpO(2) measured using the red and infrared wavelengths. After appropriate calibration, the standard deviation of the individual SpO(2)−SaO(2) differences for the two-infrared POx was smaller (3.6%) than that for the red and infrared POx (6.5%, p < 0.05). The overestimation of SpO(2) readings in hypoxemia was explained by the increase in hypoxemia of the optical pathlengths-ratio between the two wavelengths. The two-infrared POx can reduce the overestimation of SpO(2) measurement in hypoxemia and the consequent risk of occult hypoxemia, owing to its smaller increase in pathlengths-ratio in hypoxemia.
format Online
Article
Text
id pubmed-9921559
institution National Center for Biotechnology Information
language English
publishDate 2023
publisher MDPI
record_format MEDLINE/PubMed
spelling pubmed-99215592023-02-12 Overestimation of Oxygen Saturation Measured by Pulse Oximetry in Hypoxemia. Part 1: Effect of Optical Pathlengths-Ratio Increase Elron, Eyal Bromiker, Ruben Gleisner, Ori Yosef-Hai, Ohad Goldberg, Ori Nitzan, Itamar Nitzan, Meir Sensors (Basel) Article On average, arterial oxygen saturation measured by pulse oximetry (SpO(2)) is higher in hypoxemia than the true oxygen saturation measured invasively (SaO(2)), thereby increasing the risk of occult hypoxemia. In the current article, measurements of SpO(2) on 17 cyanotic newborns were performed by means of a Nellcor pulse oximeter (POx), based on light with two wavelengths in the red and infrared regions (660 and 900 nm), and by means of a novel POx, based on two wavelengths in the infrared region (761 and 820 nm). The SpO(2) readings from the two POxs showed higher values than the invasive SaO(2) readings, and the disparity increased with decreasing SaO(2). SpO(2) measured using the two infrared wavelengths showed better correlation with SaO(2) than SpO(2) measured using the red and infrared wavelengths. After appropriate calibration, the standard deviation of the individual SpO(2)−SaO(2) differences for the two-infrared POx was smaller (3.6%) than that for the red and infrared POx (6.5%, p < 0.05). The overestimation of SpO(2) readings in hypoxemia was explained by the increase in hypoxemia of the optical pathlengths-ratio between the two wavelengths. The two-infrared POx can reduce the overestimation of SpO(2) measurement in hypoxemia and the consequent risk of occult hypoxemia, owing to its smaller increase in pathlengths-ratio in hypoxemia. MDPI 2023-01-28 /pmc/articles/PMC9921559/ /pubmed/36772474 http://dx.doi.org/10.3390/s23031434 Text en © 2023 by the authors. https://creativecommons.org/licenses/by/4.0/Licensee MDPI, Basel, Switzerland. This article is an open access article distributed under the terms and conditions of the Creative Commons Attribution (CC BY) license (https://creativecommons.org/licenses/by/4.0/).
spellingShingle Article
Elron, Eyal
Bromiker, Ruben
Gleisner, Ori
Yosef-Hai, Ohad
Goldberg, Ori
Nitzan, Itamar
Nitzan, Meir
Overestimation of Oxygen Saturation Measured by Pulse Oximetry in Hypoxemia. Part 1: Effect of Optical Pathlengths-Ratio Increase
title Overestimation of Oxygen Saturation Measured by Pulse Oximetry in Hypoxemia. Part 1: Effect of Optical Pathlengths-Ratio Increase
title_full Overestimation of Oxygen Saturation Measured by Pulse Oximetry in Hypoxemia. Part 1: Effect of Optical Pathlengths-Ratio Increase
title_fullStr Overestimation of Oxygen Saturation Measured by Pulse Oximetry in Hypoxemia. Part 1: Effect of Optical Pathlengths-Ratio Increase
title_full_unstemmed Overestimation of Oxygen Saturation Measured by Pulse Oximetry in Hypoxemia. Part 1: Effect of Optical Pathlengths-Ratio Increase
title_short Overestimation of Oxygen Saturation Measured by Pulse Oximetry in Hypoxemia. Part 1: Effect of Optical Pathlengths-Ratio Increase
title_sort overestimation of oxygen saturation measured by pulse oximetry in hypoxemia. part 1: effect of optical pathlengths-ratio increase
topic Article
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC9921559/
https://www.ncbi.nlm.nih.gov/pubmed/36772474
http://dx.doi.org/10.3390/s23031434
work_keys_str_mv AT elroneyal overestimationofoxygensaturationmeasuredbypulseoximetryinhypoxemiapart1effectofopticalpathlengthsratioincrease
AT bromikerruben overestimationofoxygensaturationmeasuredbypulseoximetryinhypoxemiapart1effectofopticalpathlengthsratioincrease
AT gleisnerori overestimationofoxygensaturationmeasuredbypulseoximetryinhypoxemiapart1effectofopticalpathlengthsratioincrease
AT yosefhaiohad overestimationofoxygensaturationmeasuredbypulseoximetryinhypoxemiapart1effectofopticalpathlengthsratioincrease
AT goldbergori overestimationofoxygensaturationmeasuredbypulseoximetryinhypoxemiapart1effectofopticalpathlengthsratioincrease
AT nitzanitamar overestimationofoxygensaturationmeasuredbypulseoximetryinhypoxemiapart1effectofopticalpathlengthsratioincrease
AT nitzanmeir overestimationofoxygensaturationmeasuredbypulseoximetryinhypoxemiapart1effectofopticalpathlengthsratioincrease