Cargando…
Optimization of the dosimetric leaf gap for use in planning VMAT treatments of spine SABR cases
The dosimetric leaf gap (DLG) is a beam configuration parameter used in the Varian Eclipse treatment planning system, to model the effects of rounded MLC leaf ends. Measuring the DLG using the conventional sliding‐slit technique has been shown to be produce questionable results for some volumetric m...
Autores principales: | Middlebrook, Nigel D, Sutherland, Bess, Kairn, Tanya |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
John Wiley and Sons Inc.
2017
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC5874863/ https://www.ncbi.nlm.nih.gov/pubmed/28574219 http://dx.doi.org/10.1002/acm2.12106 |
Ejemplares similares
-
Optimizing the effective spot size and the dosimetric leaf gap of the AcurosXB algorithm for VMAT treatment planning
por: Passal, V., et al.
Publicado: (2021) -
On using the dosimetric leaf gap to model the rounded leaf ends in VMAT/RapidArc plans
por: Szpala, Stanislaw, et al.
Publicado: (2014) -
Dosimetric effects of immobilization devices on SABR for lung cancer using VMAT technique
por: Park, Jong In, et al.
Publicado: (2015) -
Evaluation of measures related to dosimetric uncertainty of VMAT plans
por: Götstedt, Julia, et al.
Publicado: (2022) -
Comprehensive dosimetric planning comparison for early‐stage, non‐small cell lung cancer with SABR: fixed‐beam IMRT versus VMAT versus TomoTherapy
por: Xhaferllari, Ilma, et al.
Publicado: (2016)